108 880 Technical/Service Manual
10.11 Analog Output Calibration
See Section 3.0, Table 3-17 for ANALOG OUTPUT parameters.
The following calibration procedure requires
a multimeter to measure voltage or current output from the analog
output module. If the option is not already installed, install it in according to the instructions included with the
option.
Note
The analog output must be calibrated after the indicator itself has been configured (Section 3.0) and calibrated
(Section 4.0).
1. Enter setup mode and go to the ALGOUT menu (see Figure 3-23):
Set OUTPUT as desired for 0-10V, 0-20mA, or 4-20mA output.
Set MIN to lowest weight value to
be tracked by the analog output
Set MAX to highest weight value to
be tracked by the analog output
2. Connect multimeter to connector J1 on the
analog output board:
For voltage output, connect voltmeter
leads to pins 3 and 4 (-V, +V)
For current output, connect ammeter leads to
pins 1 and 2 (-mA, +mA)
3. Adjust zero calibration:
Scroll to the TWZERO parameter.
Press
, 000000 will display
Check voltage or current reading on multimeter
.
Set the parameter to match t
he reading from the multimeter.
Press
or to select the digit.
Press
or to increment or decrement the value.
Press
TAR E
to move to the decimal point entry.
Press
or to adjust the decimal point placement.
Press
TAR E
to accept the displayed value.
4. Adjust span calibration:
Scroll to the TWSPAN parameter.
Press
, 000000 will display.
Set the parameter to match t
he reading from the multimeter.
Press
or to select the digit.
Press
or to increment or decrement the value.
Press
TAR E
to move to the decimal point entry.
Press
or to adjust the decimal point placement.
Press
TAR E
to accept the displayed value.
5. Verify calibration:
Return to the TWZERO/TWSPAN parameter and verify that
the calibration has not drifted.
Repeat calibration if needed.
6. Return to normal mode. Analog output function can be verified using test weights.