Table 3-12. Ethernet Menu Parameters
ETHERNET Menu
Parameter Choices Description
DHCP ON
OFF
Enables (ON) or disables (OFF) Dynami
c Host Configuration Protocal.
IPADRS 000.000.000.000
Valid IP Address
Configures the IP address for this device.
NETMSK 000.000.000.000
Valid Net Mast
Specifies the subnet mask.
DFTGWY 000.000.000.000
Valid IP Address
Default Gateway.
DNSPRI 000.000.000.000
Valid IP Address
IP address for the primary DNS server.
DNSSEC 000.000.000.000
Valid IP Address
IP address for the seco
ndary DNS server.
SERVER See SERVER Sublevel below
CLIENT See CLIENT Sublevel below
MAC 00.00.00.00.00.00
The MAC address for this device, read only.
SERVER/CLIENT Sublevel
RMOTIP 000.000.000.000
Valid IP Address
Remote IP address – IP address of the remote unit the 880 will connect to.
Client level only.
RMOTPT 1
1-65535
Remote port – TCP port number on the remote unit the 880 will connect to.
Client level only.
PORT 10001
1-65535
The TCP port number of the 880 server. Server level
only.
TRIGGE COMAND
STRIND
STRLFT
Selects the operation of the port.
•COMAND – allows operation of EDP commands and will print
•STRLFT – stream legal for trade scale data – data is transmitted at the configured
display update rate. Will also accept EDP commands.
•STRIND – stream industrial scale data – data
is transmitted at the configured A/D
sample rate. Will also accept EDP commands.
TERMIN CR/LF
CR
Termination. Selects the termination character(s) for data sent from the port.
EOLDLY 0
0-255
Port end-of-line delay, 0.1 second intervals.
TIMOUT 0
0-65535
Timeout – inactivity disconnect timeout. A connection (eithe
r client or server) is
dropped if there is no activity before the timeout expires. Time is in seconds. A
timeout value of 0 disables the inactivity disconnect.
SFMT <2><P><W7.>
<U><M><S>
<CR><LF>
Stream format – specifies the stream format
used for streaming output of scale data
(TRIGGE=STRLFT or STRIND). Alpha/numeric maximum length 200 characters.
ECHO OFF
ON
Specifies whether characters received by the port are echoed back to the sending unit.
RESPNS ON
OFF
Response – specifies whether the port transmits
replies to serial commands.
Note
If an external device (such as a printer) that may transmit unexpected
data (such as a paper low message) is connected to the indicator the
respo
nse parameter should be set to OFF to prevent a reply from the
indicator from confusing the external device.
46 880 Technical/Service Manual