
66 880 Technical/Service Manual
6.1.2 Reporting Commands
Reporting commands (see Table 6-2) send specific information to the EDP port. These commands can be used in
both setup mode and normal mode.
Table 6-2. EDP Reporting Commands
Command Function
AUDITJUMPER Returns the state of the audit jumper. A response of OK indicates the jum
per is in the ON position.
A response of “??” indicates the jumper is in the OFF position.
BUILD Returns the date and time of
the software build.
DUMPALL Returns a list of all parameter values
DUMPAUDIT Returns a list of audit trail information
DUMPCONFIG Returns a list of all parameter values except for setpoint data
DUMPETH Returns a list of all Ethernet parameter values
DUMPSP Returns a list of all setpoint parameter values
HARDWARE Returns a value that indicates which option car
d is installed in the option slot
Possible values:
000=none, 085=relay card, 153=analog out
put card, 170=CompactCom card
Example response with a relay card installed:
HARDWARE=085
VERSION Returns the 880 firmware version
DISPLAYBUILD Returns the date and time of the display module
software build. NONE is returned if no display if
connected.
DISPLAYVERSION Returns the display module softwar
e version. NONE is returned if no display if connected
P Returns the current displayed weight with units ide
ntifier, see Section 10.4
OPTVERSION Returns the software version of an
installed option card, if supported. Returns UNSUPPORTED if
not supported. Returns NOCARD if no option card is installed.
FBTEST Returns the type of Field Bus module th
at is connected to the Field Bus Option Card, if installed.
Returns NO MODULE if no module is installed. Returns NOT FOUND if no Field Bus Option Card is
installed.
6.1.3 The RESETCONFIGURATION Command
The RESETCONFIGURATION command can be used in setup mode to restore all configuration parameters to
their default values.
This command is equivalent to using the DEF
ALT function in CONFIG mode.
All load cell calibration settings are lost when the RESETCONFIGURATION command is run.
6.1.4 Parameter Setting Commands
Parameter setting commands allow the display or change of the current value for a particular configuration
parameter (Tables 6-3 through 6-22).
Current configuration parameter settings can be displayed i
n either setup mode or normal mode using the following
syntax:
command<CR>
Most parameter values can be changed in setup mode only; setpoint parameters listed in Table 6-16 can be changed
when in normal weighing mode.
Use the following command syntax wh
en changing parameter values:
command=value<CR>
where value is either a number or a parameter value. Use no spaces before or after the equal (=) sign. If an incorrect
command has been typed in, the response will be
??.
For example, to set the motion band parameter to
5, type the following:
SC.MOTBAND#1=5D<CR>
For parameters with selectable values, enter the command and equal sign followed by a question mark:
command=?<CR>
to see a list of those values. The indicator must be in setup mode to use this function.
Comentarios a estos manuales