6 880 Technical/Service Manual
1.4.6 Preset Tare (Keyed Tare)
Tare mode must be set to keyed or both for the preset tare feature to function.
1. With the scale empty and display showing zero weight, press
.
2. Display will show (000000); the focused digit will
flash.
3. To edit the value:
• Press
or to select the digit.
• Press
or to increment or decrement the value.
• Press
to move to the decimal point entry.
• Press
or to adjust the decimal point placement.
• Press
when the value is correct.
4. The display will change to the Net mode and the
PT LED lights to show the preset tare was entered.
Entering a keyed tare of zero will remove the stored tare value.
1.4.7 Print Ticket
1. Wait for the ( ) LED to light.
2. Press
to send data to a communications port.
1.4.8 Front Panel Setup
1. Press
to enter user setup mode. Use user setup to:
• View audit trail information
• Enter setup mode if audit trail is enabled
• View or set time and date
• View or clear the accumulator
• Change setpoint values and enable/disable setpoints
• View the current tare value
1.4.9 Displaying Audit Trail Information
The Audit Trail Configuration and Calibration counters can be viewed through the User Menu.
1. Press
, AUDIT is displayed.
2. Press
to display the Legally Relevant Firmware version.
3. Press
to display CALIB.
4. Press
to view the Calibration Counter.
5. Press
to return to CALIB.
6. Press
to display CFG.
7. Press
to view the Configuration Counter.
8. Press
to return to CFG.
9. Press
to return to the weigh mode.